Voltage Dip and Up Simulator
Model: VDS-2002
Conforming to IEC 61000-4-11


For additional product information, please download the VDS-2002 brochure.
VDS-2002. PDF (762 Kbyte)


Voltage Dip and Up Simulator Model:VDS-2002

FEATURES


  1. Fully compliant with all the test generator requirements of IEC standard.
  2. One unit solution for dip, swell, interruption and variation tests up to 16A 290V single phase AC
  3. Interruption test up to 16A 125V DC (*)
  4. Two motor-driven transformers approach enables switching between any voltages (*)
  5. Preset IEC test levels 0%/40%/70%/ additional 120%
  6. Two modes for 0% (interruption) test: open & short (*)
  7. Optional Windows application software available for more extensively control the unit
  8. Accurate waveforms

(*)Optional software required


SPECIFICATIONS


Items Specifications
Number of lines Single phase
Test mode Interruption
AC/DC
Local Synchronous:Short circuit
PC Synchronous: Short circuit/ Open circuit
Asynchronous: Short circuit/ Open circuit
Dip and swell Local Synchronous
PC Synchronous/ Asynchronous
Variation Local Asynchronous
(Only 2s-1s-2s standard defined test is available.)
PC Asynchronous
Input voltage range AC90~264V, 50/60Hz, DC0~125V
Output voltage range AC 0%~ 120% of input voltage, AC290V Max
DC 0V-input voltage
Output VA 4.224KVA (Continuous)
Output current capability AC 100% of input voltage: 16A rms (Continuous)
70% of input voltage: 23A rms (<5S)
40% of input voltage: 40A rms (<5S)
DC 16A (Continuous)
Peak inrush current drive capability AC 100~ 120V: >250A@(at 100% output, <10mS)
AC 220 ~240V: >500A (at 100% output, <10mS)
Load regulation 100% of input voltage 0-16A rms: <5%
70% of input voltage 0-23A rms: <7%
100% of input voltage 0-40A rms: <10%
Overshoot/ undershoot <5% (100 ohm loaded)
Rise time/ fall time 1 ~ 5 µS (100ohm loaded)
Normal voltage setting Setting by percent Local 100%
PC ~120% /10V minimum
Setting by voltage PC 10V~290V/ 5V step
Accuracy ±5V (0V~16A output)
Dip/swell level Setting by percent Local Short circuit for 0%, (Interruption) setting: 0/40/70/120% (4 steps)
PC Short/Open selectable for 0% (interruption): 0~120%
Setting by voltage PC Short/Open selectable for 0V (interruption): 0~290V(0~120%), 5 steps
Accuracy ±5V
Repetition of events (No. of events) Local 1,3, 5, 10, 30, 50,10, or continuous (8 steps)
PC 1~1,000 or continuous (1 event step)
Interval cycle Setting by cycle Local 1, 3, 5, 10, 30, 50, 100, 300, 500, cycles and 10s (10steps)
PC 0.5 ~ 5,000.5 cycles (0.5 cycle step)
Setting by time Setting for short duration PC Synchronous: 1~100s (1s step)
Asynchronous: 10ms~100s/50Hz, 8.3ms~100s/60Hz (0.1m step)
Setting for long duration PC Asynchronous: 1s ~ 10h (1s step)
Dip cycle Setting by cycle Local 0.5, 1, 5, 10, 25, 50 cycles (6 steps)
PC 0.01 ~ 5,000 cycles (0.01 cycle step)
Setting by time Setting for short duration PC Synchronous: 0.1ms~100s (0.1ms step)
Asynchronous: 0.1ms~100s (0.1ms step)
Setting for long duration PC Asynchronous: 1s ~ 10h (1s step)
Dip phase
(starting phase angle of events)
Setting by phase angle Local Synchronous: 0,45,90,135,180,225,270,315 (8 steps, 45 degrees step)
PC Synchronous: 0~359° (1° step)
Setting by time PC Synchronous: 0~19.9 ms for 50 Hz (0.1ms step)
0
~16.6 ms for 60 Hz (0.1ms step)
Voltage variation test Setting by time PC Ansynchronous:
Changing time: 0.1s~10s (0.1s step) , At least 0.1s required for 10% change of input
Changed time: 0~10s (0.1s step)
Interval: 0~100s (0.1s step)
Test level: 0~120%
Memory Local 5 tests
PC 10 steps (Test sequence stored up to 10 steps when controlled by PC.)
Equipment Input voltage AC100~240V ±10%, 50/60Hz, 120VA
External interface Optical RS-232
Dimensions 430(W) x 745(H) x 600 (D) mm (Projection excluded)
Weight Approx. 150 kG
Note: Optional software Model: 14-00036A is required for PC control.


STANDARD ACCESSORIES

  • Instruction manual

OPTIONS


  • Control software Model: 14-00036A
    • Wider parameter setting than locally allowed
    • Setting to IEC 61000-4-11 are preprogrammed
    • Intuitive setting for all test parameters
    • GUI (Graphical User Interface)
    • Sequential operation up to 10 steps
    • Test report generation
    • Upgraded software

  • RS232C Optlink set Model 07-00017A
    • Optical Modem
    • Optical Cable (5m)
  • USB Optical Module Kit Model 07-00022A
    • Main unit
    • Optical Cable (5m)
    • FD
    • Instruction manual


07-00022A Main unit

Designs and specifications are subject to change without notice.