ELECTROSTATIC DISCHARGE SIMULATOR
MODEL ESS-2000
CONFORMING TO IEC 61000-4-2

ESD Generator's brochure ESS. PDF (1568Kbyte) for additional product information.

TC-815R
Options


ESD Simulator ESS-2000 & ESD Gun TC-815R

GENERAL


Among the causes of equipment malfunction, problems caused by ESD (electrostatic discharge) are the most difficult events against which to incorporate protective measures,since the causal relation generally cannot be found easily.

Air discharge testing, which is affected by numerous variable factors, is a widely adopted test method,since the actual ESD events in the field are air discharges.

In the contact discharge method, the discharge current rises extremely quickly, approximately 1ns, independent of ESD charge voltage.  In addition, the proportional relationship between the ESD charge voltages and current peaks can be obtained, allowing the engineer to define the ESD susceptibility of the equipment under test in terms of charge voltage.  This simulator is designed to conduct both contact and air discharge testing.



FEATURES


  1. Output waveforms prescribed by IEC 61000-4-2 are available. In addition, various waveforms can be generated by changing the discharge capacitor and discharge resistor.
  2. Since the charge/discharge capacitor, resistor and discharge relay are incorporated in the gun, ESS-2000 is capable of generating current with a stable rise independent of applied voltage, making it possible to conduct a reproducible test.
  3. The four computer-controlled test modes enable the user to conduct a one touch test at each level prescribed by IEC. User may set his own program and the voltage level may also be set to change continuously.
  4. Controllable from an external computer using various communication interfaces.



SPECIFICATIONS


Item Specifications
Output voltage 0.20~30.0kV
Polarity Positive or negative
Discharge capacitor and resistor 150 PF±10%  330ohm±5% (Both changeable)
Charging resistance 10M ohm
(53M ohm for combination with TC-815R discharge gun)
Discharge mode Air discharge and contact discharge
Operation mode IEC severity level Level setting 1, 2, 3, 4
Discharge interval 0.05 ~600.0 S
No. of times of discharge 1~60000 times
Manual Discharge interval 0.05 ~600.0 S
No. of times of discharge 1~60000 times
Setting storage function Up to 10 conditions storable
Sweep Starting voltage ±0.20~30.0kV
Ending voltage ±0.20~30.0kV
Step voltage 0.00~30.0kV
Discharge interval 0.05 ~600.0 S
No. of times of discharge 1~60000 times
Setting storage function Up to 10 conditions storable
Program Voltage setting ±0.20~30.0kV
Discharge interval 0.05 ~600.0 S
No. of times of discharge 1~60000 times
No. of steps 30 steps maximum
No. of programs Up to 50 conditions storable
No. of program units Up to 100 conditions can be set.
Temp. and humidity measurement (option) Current values are displayed
Display element LCD with back light
Character display English or Japanese
Setting method Ten-key pad , Rotary control, Function keys
Auxiliary function
· Upper limit voltage setting function
· Trigger switch select function
· Auto stop function
· Inversion on the screen function
Clock function Year, month, day, hour, minute (Battery backup)
Memory function Clock work, contents of each setting and last operation display are backed up for more than 3 months with battery full charged.
External interface functions
· Warning light connecting I/F
· External trigger input I/F
· Elimination probe connecting I/F
· GP-IB I/F
Printer I/F Conforming to simple CENTRONIX I/F
Contents of print
· Currently applied voltage/ Current date/ Current temp.and humidity
· Contents of various settings
Power supply 85~264V AC          50/60 Hz
Operating temperature and humidity 0~40°C
20~75% (No dewing shall occur.)
Dimensions and weight ESS-2000 250(W) x 324(H) x 320(D)
Approx. 8.0 kG
TC-815R
(Except a cable)
75(W) x 220(H) x 210 (D)
Approx. 1.4 kG


Options




For further information, please contact us at sales@noiseken.com.