Electrostatic Discharge Simulator
ESS-2000AX CONFORMING TO IEC 61000-4-2
ESD Generator's brochure ESS. PDF (2362KB) for additional product information.

Get free "FAQ for Electro-Static Discharging Test" by answering a brief questionnaire. For more information.....
<Limited Time Offer>
Buy NoiseKens ESS-2000AX and get free "ESD Testing Manual."
Manual Sample English Chinese
TC-815R ESS-2000AX
(The Gun stand PS-806 in the photo is an optional accessory.)

GENERAL

The ESS-2000AX ESD Simulator generates highly consistent, repeatable
ESD pulses up to 30kV in both contact and air discharge modes. Compliant with IEC61000-4-2 standard, the
unit features a 5.7-inch display, a lightweight gun and optic serial
interface.
A variety of capacitor
and resistor units are available and are easily changeable within the gun. The fully- programmable, menu-driven
simulator offers three trigger modes.

FEATURES

- Fully
programmable menu-driven simulator providing three operation modes: IEC
standard, Manual and Sequence
- A
5.7-inch color LCD for easy setting and good visibility
- Unique
shape for the operator's easy access to the control and displays even when the
unit is put on the floor level (ground plane)
- Communications
with PC through the optic serial interface
- A wide
variety of the dedicated options

SPECIFICATIONS

| Item |
ESS-2000AX Specifications |
| Output
polarity |
Positive
or negative |
| Output
voltage |
0.20kV~30.0kV ±5% (30.5kV max)
0.20kV~10.0kV: 0.01V step, 10.0kV~30.0kV:0.1V step
|
| Discharge
interval |
Normal
mode: 0.05s~600.0s ±10%
Extra
mode : 1.00s~600.0s ±10%
0.05s~9.99s: 0.01s step, 10s~600.0s: 0.1s step
|
| No. of
times of discharge |
1~60,000 times preset on the counter or continuous |
| Discharge
mode |
Air/contact
discharge |
| Radiation
level |
Normal/Extra
mode |
| Trigger
mode |
Gun
trigger/main unit trigger/ external trigger |
Operation
mode |
IEC
standard |
Contact discharge: 2.0kV, 4.0kV, 6.0kV, 8.0kV
(4 voltages preset)
Air
discharge: 2.0kV, 4.0kV, 8.0kV, 15.0kV (4 voltages preset)
|
| Manual |
Contact/Air
discharge: 0.20kV`30.0kV
Sweep
(automatic voltage ramping) available
Up to
99 units (test settings) storable
|
| Sequence |
Seamless run of the selected test sequence (made of any combination of 99 units in Manual mode)
No. of
units per one test sequence: up to 30 units
No. of test sequences: up to 20
|
| Warning
lamp |
Lights
up while the HV circuitry is on
Blinks
while discharges are in progress
|
| Discharge
gun |
TC-815R |
| Charging
resistance |
10M ohm(53 M ohm for combination with TC-815R discharge gun) |
| AUX
connector |
D-SUB15
pin, female connector
Warning
lamp (optional) connection
Automatic
ESD Eliminator (optional) connection
External
interlock input
External
trigger input
|
| Optical
communication connector |
Optic RS-232C for PC communication |
| Power
supply |
100~240V AC 50/60 Hz ±10% |
| Power
consumption |
75VA |
| Operating
temperature and humidity |
+15~+35ºC, 25~75% (no condensation) |
| Character
display |
English, Chinese, Korean, Japanese |
| Dimensions and weight |
260 (W) X 230 (H) X 285 (D) mm (projection excluded)
Approx.
5.3 kg |
|

DISPLAY EXAMPLES
Display Examples in Korean Display Examples in Chinese
 |
TOP MENU |
|
 |
|
After
pressing the main switch, press the EXIT key. This places the simulator in the
initial menu, which displays three operational modes, trigger selection and
utility.
|
 |
Manual
Mode |
|
 |
If you desire to
operate the unit in the manual mode, press the corresponding function key. The parameters to be set by the operator will appear. Discharge
method (contact/air discharge), discharge voltage, number of discharges
and interval can be set. Select the parameter by pressing the
corresponding function key then the value can be altered by the Increment
or Decrement key.
Contact discharges: For contact discharge testing, after
completion of required settings, press the START button and pull the
trigger. The simulator will then
generate the required number of pulses at the required interval. Pulling the trigger again will pause the
unit. Pulling again will restart the
unit.
Air discharges: For air discharge testing, after completion
of setting, press the START key. To
carry out air discharges, pull and hold the trigger to maintain the HV relay in
the on status and approach the discharge tip to the EUT.
|
|
|
|
 |
When the SWEEP function is active, the simulator generates discharges in
an automatic voltage ramp. The initial, final and step voltages
can be freely set. In this mode, the number of discharges set is that in
each step
For example, when the simulator is set to 5kV for the initial voltage,
10kV for final voltage and 1kV for step voltage, in a way of 10 discharges
at an interval of 1 second, it produces 10 pulses at 5kV at an interval
of 1 second and proceeds to 6kV pulses, also 10 discharges. These steps continue until
the simulator has completed 10 pulses of 10kV.
Two different ways of
pulling the trigger: When the trigger
is pulled and then released quickly, the simulator operates in a way that it
pauses before it proceeds to next step voltage. For continuous operation, pull the trigger for more than 2
seconds. A sign of "CONTINUE"
is indicated on the screen and the simulator runs a complete step ramp test
without further operatorfs trigger.
|
|
|
 |
IEC
Standard Mode |
|
 |
The
four test levels are preset. Select the desired level and run a test. |
|
|
 |
Sequence
Mode |
|
 |
The simulator can store up to 99 tests including Sweep settings. Each test is given a number (called Unit No). Any combination of units selected from 99 tests in the Manual Mode can consist of one test sequence having 30 units maximum.
Twenty test sequences can be stored. For a unit setting, press EDIT button.
Settings of voltage, etc. can be done in the same way as in the Manual
mode. |

OPTIONS

 |
Automatic ESD Eliminator 01-00013A |
|
Connected to the ESS-2000AX simulator main unit, the Automatic ESD Eliminator automatically removes residual charges on the EUT. A control signal from the ESS-2000AX turns on the switch of the eliminator to drain away the residual charges to the ground.
Two operation modes are provided: turn-on after each discharge and turn-on
after preset number of discharges (Not compliant with the standard).
|
 |
|
|
 |
Gun holder 03-00056A
|
|
A gun
holder that can be screwed to the left side of ESS-2000AX.
|
|
|
|
03-00056A
Usage example
|
 |
Warning lamp 11-00014A |
|
|
 |
AUX
connector junction box 05-00052A |
|
Warning
Lamp and Automatic ESD Eliminator options and external trigger circuit can be simultaneously
connected. |

|
 |
Optical USB module 07-00022A |
|
|
Optical conversion adapter to be free from the EM interference when communicating with PC. |

|
 |
Optical RS232 module 07-00017A |
|
|
Optical
conversion adapter to be free from the EM interference when communicating with
PC |
|
|
|
|