株式会社ノイズ研究所

I am a user of the Noiseken Discharge Gun GT-30RA (and either of ESS-S3011A/B3011A/1611A ESD simulator main unit). Does the use of capacitor-resistor unit of 100pF/1.5K ohm and 200pF/0 ohm make my ESD system compliant to component level ESD test standards for human body model and machine model, respectively, such as

NoiseKen

1) AEC-Q100-002, ESDA ANSI/EOS/ESD-STM5.1, IEC 61340-3-1, IEC 60749-26, JEDEC JESD22-A114, JEITA EIAJ ED-4701/300 Test method 304 or MIL-STD-883 Method 3015.7, and

2)AEC-Q100-003, ESDA ANSI/ESD STM5.2, IEC 61340-3-2, IEC 60749-27, JEDEC JESD22-A115 or JEITA EIAJ ED-4701/300?

No. The test waveform specifications for component level ESD test standards are very different from the one specified in the IEC 61000-4-2 standard. For component level ESD testing, two products ESS-6002 (2kV) and ESS-6008 (8kV) are available.