株式会社ノイズ研究所

What test instrument and accessories are required to measure the output waveform from an ESD generator?

NoiseKen
The following equipment and accessories are required:
06-00067A ESD current target
00-00011A 20dB attenuator
02-000132A coaxial cable
00-00010A 6dB attenuator (1 piece or 2 pieces depending on test voltage
02-00133A coaxial conversion SMA to BNC, when your scope input is a BNC type)
Oscilloscope with >2GHz bandwidth for IEC 61000-4-2 or >1GHz bandwidth for ISO 10605
The ESD current target shall be mounted on a metal plate working as a shielding wall (03-00052B 1.2 x 1.2 m, specifically designed for this purpose, also available) and the scope shall be placed to the output side of the target to be electromagnetically separated from the ESD generator.
All the said items shall be connected in the following order to form an ESD target-attenuator-calibration chain: 06-00067A ESD current target -00-00011A 20dBattenuator – 02-000132A coaxial cable – 00-00010A 6dB attenuator (1 piece or 2 pieces depending on test voltage)
The following options make easy for you to create the conditions required in the IEC/ISO standards.
03-00060A Cable Positioner
The discharge return cable shall be connected to the plane onto which an ESD current target is installed at a distance of 0.5m below the ESD current target and shall be pulled backwards at its midpoint to form an isosceles triangle.
03-00061B Discharge Gun Mount
This mount is attached to the 03-00052B plane and holds the discharge gun.

News

  1. 07/25/2024 Information

    Summer Holiday Notice

  2. 07/03/2024 Information

    NoiseKen will be exhibiting at “Techno-Frontier 2024, the 37th EMC and Noise Countermeasures Technology Exhibition”.

  3. 04/25/2024 Information

    Long holiday information

  4. 03/15/2024 Products & Services

    New release of Impulse Noise Simulator INS-S100

view all

Contact