Impulse Noise Simulator INS-S420
To Solve the Real Trouble in the Market
Noise simulator can simulate high frequency noise generated ON/OFF at contact point of switch or relay, arc caused by electric motor. It can evaluate the resistibility of electric devices.
Pulse contains high frequency and by energy volume is changable by adjusting pulse width. The high reproduction noise test of noise trouble in the market can be conducted.
- Button touch instead of coaxial cable replacing reduces pulse width setting time.
- Pulse waveform stability is improved, so high reproduction test is available.
- Cost is cut down because consuamable parts are reduced.
- Common mode/normal mode test is easily to switch by short plug.
- “Test time setting” new function simplifies test time setting.
- Wiring becomes easier because 50Ω resisitor is built in simulator.
- Repetition cycle becomes faster. Due to high repetition, mal-function occurrence rate is up and test time is to be shortened.(Only INS-S220)
- AC plug of EUT can be inserted directly by outlet panel.(option)
- Various tests are available by using different probes and coupling clamp.(option)
- EUT test with 3 phase is available by external CDN.(option)
- Dedicated software simplifies testing with various test conditions (Option. Applicable to INS-S420 only)
To solve the trouble in the market high frequency, energy volum of test pulse can be ajustable
Even narrow pulse with 50ns-100ns width contains less energy, twice fast transient due to rise and fall and inducted coupling occurred by sharp electromagnetic field effect electric circuit greatly.
Wide pulse with 800ns-1000ns contains more energy, so voltage fluctuation is easily to effect circuit.
The rise time of impulse simulator is faster than IEC61000-4-4 fast transient/burst test, so spectrum is high. When it injects noise to EUT, noise is easier to invade electric circuit internally.
Spectrum and amplitude is different due to Impulse width, so it is recommended to test with different pulse width.
.
Setting is simplified. Setting time is shortened.
It is troubesome to change the special coxial cables manually in old way. Setting time and connection mistake can be reduced because setting can be operated by button. requires for manually switch dedicated coaxial cables.
Cost is cut down. Consumable parts are reduced
We adopt semiconductor type relay instead of the mercuray relay in old type.
Also, cost on consumable exchange is reduced because pulse width setting cable (consumable) is no longer needed.
Connection is simplified. Connection time is shortened.
Outlet panel to which EUT is easyily to be connected is adopted.
EUT is easy to connect by using outlet panel (option) complying to each country's socket shape.

Noise countermeasurement is easy. The malfunction generating position can be identified.
From power supply line, signal line, harness, enclosure to PCB, various noise injection options are ready. The malfunction generating position is easy to identify.
Specification
Parameter | Specification | ||||||||
Pulse setting range | Output voltage : 0.50kV 〜 4.00 kV ±10% 0.01kV step Pulse width : 50ns ±15%, 100ns〜 1000 ns ±10% 50ns step Variable : 10ms 〜 999 ms ±10% 1ms step |
||||||||
Output voltage | 0.5〜4.00kV±10% (10V step) | ||||||||
Polarity | Positive / Negative | ||||||||
Rise time | 3ns Max | ||||||||
Putput impedance | 50Ω | ||||||||
Termination resistance | 50Ω | ||||||||
Pulse repetition mode |
|
||||||||
Memory capacity | 5 tests | ||||||||
Operating time | 1s 〜 999s ±10% 1s step | ||||||||
Superposition phase switching | L(+), N(−), PE / PULSE OUT ※Manual switching by coaxial cables | ||||||||
Superposition mode | Common mode / Normal mode ※Manual switching by short plugs | ||||||||
EUT power capacity | Single phase AC240V / DC125V 16A (L(+), N(−), PE) | ||||||||
External control | RS-232C compliant optical communication | ||||||||
Power source/Power consumption | AC100 〜 240V 50Hz/60Hz | ||||||||
Operating temperature/Humidity range | 15 to 35℃/25〜75% | ||||||||
Dimension/Mass |
(W)430×(H)349×(D)540mm(Does not include protrusions)/Approx. 23kg | ||||||||
High voltage coaxial connector | NMHV NoiseKen custom | ||||||||
Accessories | Coaxial cables 30cm(02-00013A):2, SG setting short plugs(02-00106A):1, SG cables(05-00103A):1, Outlet Panels:1, AC cables:1, Instruction manuals:1, Accessories bags:1 |